PS4-MT3-N-1/R-1 nitrogen purifier, 110V. This MonoTorr purifying alloy is selected to purify nitrogen; others will purify rare gases (H3, Ne, Ar, Kr, Xe). The specifications below are for this nitrogen Phase ll, 3000 series 110 VAC purifier.

Click here to see the serial number back plate.
The purifier has an electropolished interior surface, minimal dead space, and is easy to install. The purifier is completely assembled, with 1/4" male and female ultra-high purity (UHP) face seal fittings for gas inlet and outlet. It includesinlet and outlet stainless steel diaphragm valves, 0.003 micron particle filter, and an on-board electronic control unit. All nitrogen and rare gas purifiers have a Life Status Sensor (LSS) as an impurity breakthrough indicator.
All impurities are chemically bonded to the purifier material. The purifier will not release the trapped impurities under vacuum, high pressure, room temperature or maximum design temperature, even when purifier nears its maximum capacity for impurities.
Read more specifications at SAES Getters Pure Gas Technologies.
| Specifications | Rare Gas & Nitrogen |
|---|---|
| Maximum Flow Rate | MT3: 50 SLPM (1.0 SLPM minimum) MT15: 75 SLPM (1.5 SLPM minimum) |
| Nominal Flow | MT3: 5 SLPM, MT15: 15 SLPM |
| Maximum allowable working pressure | 10 bar (150 psig) |
| Operating temperature | Nitrogen: 350°C, Rare Gas: 400°C |
| Outlet gas temperature | 50°C |
| Installed power | MT3: 300W Optional 120 VAC: MT15: 600W MT3: 330W Optional 220 VAC: MT15: 660W |
| Getter confinement | 15 um in/out porous SS filter |
| Inlet connection | 1/4" face seal female fitting |
| Outlet connection | 1/4" face seal male fitting |
| Pneumatics connection | 1/4" NPT female fitting |
| Valves | Three, diaphragm type, air actuated |
| Pneumatics utilities | Nitrogen or clean dry air at 80-100 psig, filtered to 10 um |
| Filter | .003 um all-metal type |
| Control unit | Discrete analog temperature controller with alarm and shut-down features |