OEM Rebuilt MKS 651C Pressure Controller
Pressure Controller: 651C Self-Tuning/Digital PID
$1,500 OEM Rebuilt 90 day warranty. The MKS Instruments 651C self-tuning pressure controller provides a fast response to set point under varying operating conditions when operating Type 253 or 653 Valves. Includes integral power supply for valve and most general purpose transducers.
Model#: 651CD2S1N
MKS Instruments 651 Serial RS-232 Connector Specifications
NOTE:
The MKS RS-232 Serial communications cable (CB651-10-10) must be used for the 9 pin Serial Interface connector. Unlisted pins on this connector are not compatible with the Type "D" 9 pin, industry standard cable.
Pin Description
- No Connection
- Transmit Data
- Receive Data
- No Connection
- Digital Ground
- Reserved
- Reserved
- No Connection
- No Connection
I/O Connector
- PLO relay #1 - NC contact
- PLO relay #1 - NO contact
- PLO relay #2 - NC contact
- Digital Ground
- Learn system (Low)
- Hold both pins 6 and 11 Low to select analog set point with position control
- Softstart (Low)
- Close valve (Low)
- Reserved
- Analog set point divided by 10
- Hold only pin 11 low to select analog set point with pressure control; Hold both pin 6 and 11 low to select analog set point with position control
- 1Select set point E (Low)
- Select set point D (Low)
- Select set point C (Low)
- Select set point B (Low)
- Select set point A (Low)
- Reserved
- Reserved
- Valve open status (Hi=open)
- PLO relay 1 - common contact
- PLO relay 2 - common contact
- PLO relay 2 - NO contact
- Valve closed status (Hi=closed)
- Reserved
- Remote zero (Low)
- Stop valve (Low)
- Open valve (Low)
- PLO #2 status (Low=out of limit)
- PLO #1 status (Low=out of limit)
- +15 V Output
- -15 V Output
- Power Ground
- + Set point Input
- - Set point Input
- Analog Ground
- Pressure output voltage
- Position output voltage
Transducer Connector
- Pin Description
- + 15 V Supply
- + Pressure Input
- Reserved
- Reserved
- Power Ground
- - 15 V Supply
- +15 V Supply
- Reserved
- -15 V Supply
- Reserved
- Digital Ground
- - Pressure Input
- Reserved
- Reserved
- Chassis Ground