Rebuilt Alcatel ASM Leak Detectors

ASM 10
ModelPrice
ASM 10$7,500
ASM 110 Turbo CLask
ASM 120Hask
ASM 121 hask
ASM 180 task
ASM 180 tdask
ASM 181 td2ask

ASM 110 Specifications
Maximum operating pressure of the analysis cell10-4 mbar
Pumping speed for He
in front of the analysis cell
30 l/s
at the inlet valve8 l/s
Intrinsic sensitivity of the analysis cell3 , 5.10-4
Variation of sensitivity as a function of pressure between 10-6 and 10-4 mbar±40% maximum
smallest detectable signal2.10-11 atm.cm3/s He
Leak rate detection range:
      direct2.10-11 to 1.10-5 atm.cm3/s He
      closed loop circuit1.10-5 to 3.10-1 atm.cm3/s He
Pirani gauge detection1 to 100 atm.cm3/s He
Response time
      scale 10.10-101.5 s
      scale 3.10-91 s
      scale 10.10-9 - 10.10-60 , 5 s
Filament protection (triode)off at 2.10-4 mbar
Turbomolecular pump protection (Pirani gauge)Closing of the safety inlet valve at 5.10-2 mbar
Audio signaladjustable on the whole range
Air roughing pumping speed at inlet port1 l/s