Veeco MS-40 Leak Detector

The former Veeco Instruments family of leak detectors is now manufactured, sold and serviced by Vacuum Instrument Corporation (VIC). It is also sold as an Edwards Spectron 600.
The MS-40 combines high test port pressure with a large capacity internal mechanical pump. The MS-40 utilizes a 180° deflection dual magnetic sector mass spectrometer tube with built in high vacuum ion gauge. Two long life, low current filaments are provided. A self cleaning heated repeller and automatic closed loop emission control are at the heart of the MS-40 leak detector design. Heavy ion traps reduce the need for deionization of the mass spectrometer tube by an order of magnitude. The mass spectrometer is sensitive to Helium Mass 3 or Mass 4 and is operator selectable.
The ability to perform direct flow leak detection allows for orders of magnitude cleaner leak detection than reverse flow only leak detectors. For applications which don't require 100% dry leak detection, the ability to do direct flow only leak detection with the conventional MS-40 can offer an alternative to purchasing a dry leak detector.
Features and Benefits:
- Simple one-button start up. The MS40 system is ready in less than three minutes with zero operator intervention.
- Fully automatic operation, tuning and calibration. Reduces operator training, making the MS40 the easiest to use and most automated portable in the industry
- Three modes of operation in a single unit. Direct mode can be used for higher sensitivity. Reverse mode is available for faster cycle time. Combination mode for automatic switching between reverse and direct flow is also available to the operator. Built in test diagnostics display system error codes and isolate faults for trouble-free maintenance and quick repair.
- Mini-sniffer probe port. Allows simple sniffing operation through a dedicated sniffer port connection.
- Turbo pump with liquid lubricated ceramic bearing. Provides the ability to perform preventative maintenance. Greatly increases turbo life. Less heat and better lubrication than grease lubricated bearing.
- Solenoid valves designed specifically for vacuum use. Valves last up to 20 million cycles.
- Dual magnetic sector mass spectrometer with resolving power of 14. Lower background levels during testing. Leak detector does not mistake adjacent peaks for helium.
- The industry's only de-scan switch allows for positive identification of helium at any time during testing.
- RS-232 port provides for external control. Ideal for integrating to automated control systems and personal computers.
Performance:
- Sensitivity: 4.0 x 10-11 ATM-cc/sec in direct mode* 4.0 x 10-11 ATM-cc/sec in direct mode* 4.0 x 10-10 ATM-cc/sec in reverse mode* 6.0 x 10-9 ATM-cc/sec in reverse mode*
- Response Time in direct Mode: less than 2 seconds* (*Resolution resolving power of 14 at mass 4)
- Response Time in Reverse Mode: less than 3 seconds*
Leak Range:
- 10.0 x 10-5 ATM-cc/sec to 4.0 x 10-11 ATM-cc/sec
- 10.0 ATM-cc/sec to 4 x 10-11 ATM-cc/sec w/optional dedicated external roughing pump
Maximum Inlet Pressure Test Mode Atmosphere Gross (with dedicated external roughing pump ) 7.5 Torr Fine
SYSTEM:
- Startup: Less than 3 minutes
- Time to test: Less than 1.5 seconds on a blank test port.
- Tuning and Calibration: Automatic
- Ranging: Automatic or Manual
- Quick Calibration Check: Automatic
Flow Modes: The MS-40 uses Direct, Reverse or Combination Flow Modes. The leak detector may be set to automatically switch between modes based on leak rate and test port pressure, or be locked in either Direct or Reverse flow testing mode.
Dedicated Sniff Mode: One button allows simple sniffing operation.
Selectable Measurement Units: Pressure: Leak Rate
- mtorr STD-cc/sec
- mbar-1/sec
Vacuum System:
- Turbomolecular Pump: 52 l/s turbomolecular pump with ceramic bearing.
- Mechanical Foreline/Roughing Pump: Built-in 1.5 CFM direct drive foreline/roughing pump.
- Optional 7 CFM or 16 CFM auxiliary roughing pump available.
- Valves: Reliable, fast acting solenoid valves rated for millions of cycles.
- Easy to clean without special tools, designed for long, trouble-free life.
- Vacuum Gauging: Two Pirani gauges monitor foreline and test port pressures.
- An ionization gauge is built into the spectrometer tube ion source for reduced maintenance.
- A built-in, maintenance-free HiVac gauge utilizes existing focusing plates within the source and eliminates the need for a high maintenance cold cathode discharge gauge.
- Access: Hinged doors and pop-off side panels for serviceability of pump oil, air filters and internal calibrator.
- Calibrator: Internal temperature compensated Helium leak standard, Traceable to NIST with quick disconnect for easy recalibration.
Physical Data:
- Size: 15"(H) x 20"(W) x 14"(D) (38.1 cm (H) x 52.7cm (W) x 35.6 cm (D))
- MS-40 Weight: 87 lbs. (39.5 kg.)
- MS-40 DRY Weight: 96 lbs. (43.6 kg.)
Power Requirements: 115V 60Hz. 8A 230V / 240V 50Hz. 5A