The Turbo-V controllers are frequency converters with self diagnostic and protection features. The half rack unit has a multifunction alphanumeric display for pump status and error code diagnostics. It displays rotational speed as the pump starts up and indicates when full speed is reached. At any time during the operation of the pump, the speed, current, power, and bearing temperature can be displayed. Its microprocessor acts as a pump cycle log, and can display the number of vacuum cycles, the cycle time for the current cycle, and the total operating hours on the pump.
The Navigator series, NAV on the tag, mounts directly onto the bottom of the turbo pump. It has no display since it is meant to be linked into your system's process control.
| Specification | 150HT | 300HT | 550 | 700HT | 1000HT | 1800 | 3500 | 6000 |
|---|---|---|---|---|---|---|---|---|
| Input | 120 / 220 V 1Φ | |||||||
| Maximum Input power | 350VA | 600VA | 700VA | 1300VA | 3000VA | |||
| Output Frequency | 1033Hz | 933Hz | 700Hz | 633Hz | 450Hz | 250Hz | 233Hz | |
| Maximum output power | 200W | 250W | 325W | 350W | 450W | 680W | 1100W | 1500W |
| Operating temperature | 0° to +40°C | |||||||